Auflistung i-com Band 15 (2016) Heft 3 nach Schlagwort "Configurable Footswitch"
1 - 1 von 1
Treffer pro Seite
Sortieroptionen
- KonferenzbeitragA Configurable Footswitch Unit for the Open Networked Neurosurgical OR – Development, Evaluation and Future Perspectives(i-com: Vol. 15, No. 3, 2016) Dell’Anna, Jasmin; Janß, Armin; Clusmann, Hans; Radermacher, KlausFootswitches are used in the neurosurgical operating room for human-device-communication every day. However, problems, such as shifting or confusion of footswitches, often occur due to the parallel usage of up to 5 device-specific footswitches, resulting in a significant burden for the surgeon. There are no footswitches available which offer an optional central activation of different devices from various manufacturers and a reconfiguration during usage. Therefore, a new concept of a configurable central footswitch unit has been developed for optional activation of different devices in an open networked neurosurgical OR setting. In a user-centered evaluation 9 surgeons used both, the configurable central footswitch unit and 4 device-specific footswitches, for a cross-over experiment in an experimental OR setting. It shows that all surgeons were able to handle the configurable footswitch autonomously and that efficiency in surgeon-device-communication can be increased.